调节固定器与膝功能架的生物力学测试
摘要点击次数: 1595   全文下载次数: 1309       
作者Author单位AddressE-Mail
师宜健 Shi Yi-jian 天津医院骨科研究所 300211 Institute of Orthopaedics, Tianjin Hospital 300211  
韩慧 天津医院骨科研究所 300211 Institute of Orthopaedics, Tianjin Hospital 300211  
邢晖 天津医院骨科研究所 300211 Institute of Orthopaedics, Tianjin Hospital 300211  
王志彬 天津医院骨科研究所 300211 Institute of Orthopaedics, Tianjin Hospital 300211  
朱式仪 天津医院骨科研究所 300211 Institute of Orthopaedics, Tianjin Hospital 300211  
顾云伍 天津医院骨科研究所 300211 Institute of Orthopaedics, Tianjin Hospital 300211  
期刊信息:《中国骨伤》1993年,第6卷,第4期,第9-10页
DOI:doi:10.3969/j.issn.1003-0034.yyyy.nn.zzz
基金项目:
中文摘要:津医院创伤骨科设计的调节固定器与膝功能架治疗股骨髁间骨折,取得了较好的临床效果。本文报导对该固定器进行临床生物学测试分析,结果表明:外内侧支撑杆的牵引力分别为5.5kg和6.8kg,压垫下压力约为4.5kg,动态测试结果表明:所测试的各部分力值均有约10~20%的波动,符合骨折固定的弹性理论,即相对固定有利于骨折的愈合。
【关键词】骨折固定器  股骨髁间骨折  生物力学  实验研究
 
Biomechanical tests on the adjustment fixator and knee joint functional frame
ABSTRACT  Better clinical results had been obtained with the adjustment fixator and knee joint functional frame in treating intercondylar fracture of the femur designed by the Deparment of Traumatology, Tianjin Hospital. In this Paper, clinical biomechanical measure ments were made, the results indicated that the traction force of the external and internal supporter were 5. 5kg and 6. 8kg respectively, the pressure force underneath the pressing cushion was about 4. 5kg. Results of the motive measurement indicated that the pressure value of all parts bore a fluct uation around 10-20%. It accorded with the theory of elastic fixation of fracture treatment.
KEY WORDS  Fracture fixator Biomechanics Experimental study
 
引用本文,请按以下格式著录参考文献:
中文格式:师宜健,韩慧,邢晖,王志彬,朱式仪,顾云伍.调节固定器与膝功能架的生物力学测试[J].中国骨伤,1993,6(4):9~10
英文格式:Shi Yi-jian.Biomechanical tests on the adjustment fixator and knee joint functional frame[J].zhongguo gu shang / China J Orthop Trauma ,1993,6(4):9~10
阅读全文  下载  查看/发表评论  下载PDF阅读器
关闭




版权所有:《中国骨伤》杂志社京ICP备12048066号-2  版权声明
地址:北京市东直门内南小街甲16号,100700
电话:010-64089487 传真:010-64089792 Email:zggszz@sina.com

京公网安备 11010102004237号